7.1.3 Connecting the service PC to the BV family system The service PC can be connected with a cross cable to the BV Family system. In case of integrated DICOM functionality: contact the network administrator for an IP address for the service PC.(Usually this can be done by selecting: Programs > PMS Service Work Bench > Security Toolbox.Start the Security toolbox to check this.A connected hardware key level 0 for X-ray or IST level 0 account with known password.Windows 2000 or XP with administrator rights.7.1.2 Requirements of the service PC The requirements for the service PC are: WARNING Together with a restart of the system, always close BV-Scope and restart it. WARNING After adjustments of the system always restart the system to ensure, the new adjustments will be activated. Next, the BV-Scope menu items that are available to BV-Scope administrators (with authorization level "A") are described. would need to know software version to know for sure 7.1 Service PC 7.1.1 Introduction This chapter describes how to configure a connection between the service PC and the BV Family system. ![]() if you do not have enable service you have a very old system that requires a parallel dongle most likely. Per the manual I have it does not say db9 to db25 it says network crossover.if there was an email address I would send you the manual. ![]() Re: re: re: Philips Libra X-Scope Connection